Turbine Flowmeter Faults and Trouble-Shoot Methods
Turbine Flowmeter Faults and Trouble-Shoot Methods
For faults and trouble-shoot methods, please see Table as below:
Faults | Possible causes | Trouble-shoot methods |
The liquid flows regularly,but the equipment fails to provide indication and increase of quantity | 1)Broken supply circuit or signal circuit or bad contact | 1)Use a universal instrument to check the circuits and remove the breakdown point. |
2)Failures or bad contacts of PCB and connectors | 2) replacing PCB | |
3)Use a iron bar to move quickly under the detecting head, in the case of no signal output, check whether there is line cut and sealing-off at the soldered dot in the coil. | ||
3)Breakdown of pre-amplifier | ||
4)Voltage supplied for the pre-amplifier is too low | 4)Rise source voltage up to the required level. | |
5)The impeller is clamped down and cannot move. | 5)Remove the foreign matter and clean or replace the damaged parts,and after replacing the parts, the transducer should be recalibrated. | |
1)The filter is blocked up and the pressure loss is growing bigger to reduce flow. | 1)Remove foreign matters in the filter. | |
2)The valve plug on the pipeline is loosening, and the valve opening is reduced. | 2)Repair or replace the valve. | |
3)The impeller is blocked up by impurity or foreign matters enter.The clearance between bearings,which bring about bigger resistance, thus reducingthe speed. | 3)Clean the flow meter, if necessary,recalibrate it. | |
Flow indication is lowering gradually.When the fluid is not flowing,but the indication is not zero, and the Indication value is not stable. | 1) Bad shielding and groundingof the transmission line,Interference of outside,electromagnetic field. | 1)Check the grounding and eliminate the interference. |
2)Pipeline vibration causes shake of impeller. | 2)Fasten the pipeline or add supports in front and back of the flow meter. | |
3)Leakage of globe valve | 3)Repair or replace the valve. | |
4)Circuit or electronic components in the display Instrument are damaged and bring about interference. | 4)Take “short-circuit” method or check them one by one to find out the breakdown point. | |
The indicated flow is not in conformity with the actual flow. | 1)The impeller as been corroded and the blades are deforming. | 1) Repair the impeller or replace it and then recalibrate the transducer. |
2)Impurities have stopped the impeller from turning. | 2)Remove such impurities. | |
3) The output signal of detecting coil is not normal. | 3) Check the insulation resistance and on resistance of the coil. | |
4) Fluid temperature is too high or too low, which results in big change of clearance between bearing and shaft.
5) Inadequate back pressure toresult in cavitation.
6) Fluid viscosity is becoming bigger as a result of temperature influence.
7) Reverse flow as a result of failure to install a clack valve. | Take tailor-made measures to remove the fault. | |
8) Leakage of the bypass valve | 8) Close the bypass valve, if necessary, replace it. | |
9) Distortion of flow velocity distribution in the upper streamof the transducer or a pulsating flow. | 9) Try to find out the cause of such distortion or pulsating flow and take appropriate measures to eliminate such influence. | |
10) Display instrument fault | 10) Repair the display instrument | |
11)Incorrect connection of display instrument | 11) Try to correct the connection | |
12) Wrong setting of display instrument | 12) Try to correct the setting | |
13) The actual flow surpass the specified flow range. | 13) Replace the flow meter with one of proper diameter. |
The user must contact us in case of failure of transducer during operation. In that case, please make detailed description of the instrument fault, operating conditions, measured fluid characters, model and serial number and other details of the faulted instrument.When it is necessary, please attach the schematic diagram of the installed pipeline as well as major technical parameter, for purpose of our better service for you.
Related Industry Knowledge
- Vortex Flowmeter Design & installat...
- Vortex Flowmeter Designing & instal...
- How to choose suitable Vortex Flowm...
- Vortex Flowmeter technical indexes
- Vortex Flowmeter working Principle
- Mini oval gear Flowmeter ordering code
- Micro Metal tube Flowmeter Feature:
- Metal tube Rotameter Ordering code
- Turbine Flowmeter Guide for order
- Turbine Flowmeter Caution matters f...
- Turbine Flowmeter Technical paramet...
- Turbine flowmeter Wiring
- Turbine flowmeter Flange connection...
- Turbine flowmeter Thread connection
- Specification Codes Turbine Flow meter
- Turbine flowmeter function
- Micro flow range Metal tube Rotameter
- Turbine flowmeter Caution Matters
- Digital Pressure gauge of ordering